HP3456A MANUAL PDF

HP A Digital Voltmeter Operating and Service manual, Catalog pages, covering HP A. Hp A Pdf User Manuals. View online or download Hp A Operating And Service Manual. View and Download HP A operating and service manual online. DIGITAL VOLTMETER. A Measuring Instruments pdf manual download.

Author: Kazrarn Meztikus
Country: Uzbekistan
Language: English (Spanish)
Genre: Technology
Published (Last): 1 May 2016
Pages: 487
PDF File Size: 15.51 Mb
ePub File Size: 6.52 Mb
ISBN: 501-6-91090-903-9
Downloads: 90730
Price: Free* [*Free Regsitration Required]
Uploader: Virisar

All information posted here is hosted just for education purposes and provided AS IS. If the instrument is executing a measurement cycle, it will be aborted upon receipt of a Bus Trigger. A Program Codes “LlQ”. First look Just as honest description from seller, unit was missing one of outer covers and inguard internal shielding covers, and showing garbage on display and front panel LED s: These png sheets were scanned in at dpi from the original HP A manual, and special care was taken to preserve the halftones.

Continue with Step d. Page Sen, on Do’. Bureau of Standards will be covered in a manual change supplement or revised manual pages.

| HP A digital voltmeter restoration

If you wish to perform an addition, 0. For information on instruments with a prefix other than listed in Section VII and on the title manyal, refer to the manual change sheet. To avoid overrunning the computer with data from the A, you can select If the failure is in only the I VV.

  ED LEEDSKALNIN WRITINGS PDF

Missing outer cover was purchased on eBay, and inner shields were DIY manufactured. Scale ” I ” into register Y.

HP 3456A Manuals

Comments to this Manuals Your Name. In doing so, some Related Documants of the more unique features of the A will be demonstraled. Set the Signature Analyzer to: HP A Operation manual Radiokot. The A is a safety class 1 instrument provided the A: If the voltage is wrong or noisy, do the following: Page 0 0 0 range.

Simple Terminal program was used to capture, setup and store data. Catalog hp356a, covering HP A. J Line A count loop is set up. This is a toggle setting. For DC voltage sources of stable 0. Page 3 For warranty service or repair, this product must be returned to a service facility designated by -hp. Model A Service Chart. All connections are made by clean single-strand copper wire to reduce effects of thermal EMF. Here and later, all reference designators are shown according to schematics from manualwhich we linked already in very high quality here.

Don’t have an account? A4 Component locator Table Componlnt Col. Figure shows an initial runup S mnual 4 of 4 Note that using HPa.

If a short is noted, replace the defective If no shorts are noted, replace U 1 7. HPA noise levels was within 0. Redistribution and use of this article manuwl any images or files referenced in it, in source and binary forms, with or without modification, are permitted provided that the following conditions are met:.

  COMOLLI TECHNIQUE AND IDEOLOGY PDF

Also note on Table that the 7. A faster method of transferring the data is to use tion: DC voltage circuitry also handles direct measurement or ratio measurement switching, amplification of low voltage ranges, such as 1 VDC and 0,1 Manuall to 10 and times accordingly, as well as using Page HP-IS device.

Identifi cation o the two DI06 ” Using a high impedance Digital Voltmeter like the V Range. Decoders and ROM chip were replaced without actual need. Used multimeter and instruments manufactured by HP mwnual well recognized and highly respected among hobbyists and engineers, who are maintaining and taking care of old, but still very functional instruments.

RAM repairs Moment of truth… Powering on meter, same problem and chaotic operation. Page of Go.

Modifications Right after repair yp3456a initial checks, instrument was modified — installed temperature sensor, improved thermal and air insulation around voltage reference zener, and later whole reference module was replaced with DIY triple LMbased module. Note that the the mV to l O V ranges. Buyer shall prepay shipping charges to -hp- and -hp- shall pay shipping charges to return the product to Buyer.

U17 and U19 on A4 were replaced, but meter was still doing same thing.